Determination Of Urinary Microalbumin Reagent

[Sample requirements]
Urine samples must be collected in a clean, dry plastic or glass container. Urine is best used in the morning. Urine can also be tested 24 hours or 12 hours or overnight after urine collection, and urine does not need any special treatment or prior treatment. Urine in 2-8 ℃ storage can save 3 days. The frozen urine must be restored to room temperature and stirred well before testing.

[Test method]
Read the instructions carefully before use.
Restore kits, urine samples and quality control products to room temperature before testing.
Remove the kit from the sealed bag and mark the sample number on the kit.
Follow the instructions and add 2-3 drops of urine to the injection hole.
Wait for the result, read the resul 5 minutes after adding the sample. Don't read the results after 10 minutes.

[Reference range]
The minimum detection amount of this product for MAU is 15-20mg /L, and the test result >20mg/L is positive, while <15mg/L is negative.

[Interpretation of test results]
The result is judged by comparing the color intensity of C and T lines. Results were classified into three grades: negative (<15mg/L), threshold range (15-20mg /L), positive (>20mg/L).

Negative: there is a red line at the position of detection line (T line) and quality control line (C line), and the color of T line is darker than that of C line. It indicates that the concentration of albumin in the sample is within the normal reference range, and the concentration is <15mg/L.

Threshold range: a red line appears at the detection line (T line) and quality control line (C line) of the result observation window, and the color intensity of T line and C line is similar. The concentration of albumin in the sample was 15-20mg /L.

Positive: a red line appears at the position of the detection line (T line) and the quality control line (C line) of the result observation window, and the T line is lighter than the C line. The concentration of albumin in the sample was >20mg/L. The red line in the substance control line (C line) indicates that the concentration of albumin in the sample is >200mg/L.

No effect: there is no red line in the position of quality control line, indicating that the result is invalid and should be retried.

Chemtron biotech mau test is of high quality. Learn more about the mau test from chemtron biotech.
Related News
  • Determination of FSH

    Determination of FSH

    August 8, 2018Determination of follicular stimulating hormone (FSH),[normal reference value]Male: 2.5-15u /L.Female: 4-20u /L; Postmenopausal: > 40-20ou /L.The normal reference value is 2.5-15u /L for men and 4-...view
  • Perimenopausal Syndrome

    Perimenopausal Syndrome

    July 26, 2018Perimenopausal syndrome, also known as menopause syndrome (MPS), refers to a series of disorders of autonomic nervous system mainly caused by sexual hormone fluctuation or reduction before and after m...view
  • Classification of Fob Rapid Test

    Classification of Fob Rapid Test

    July 26, 2018Fecal occult blood test paper can be divided into two types according to the application method: traditional method (self-determination of feces) and new method (direct detection without feces).3.1 tr...view
  • Application And Sample Requirements Of Ovulation Test Paper

    Application And Sample Requirements Of Ovulation Test Paper

    August 8, 2018Intended useThis reagent was used to qualitatively detect the changes of luteinizing hormone in women's urine in vitro, so as to determine the "safe period" of ovulation time and women's menstrual cyc...view
  • Characteristics of HCV Detection Kit

    Characteristics of HCV Detection Kit

    July 26, 2018It shortens the "window period" of HCV infection, which is 70 days to 2 years shorter than the "window period" of HCV antibody detection currently used in the marketThe vaccine can be used to diagnose...view
  • Significance Of Determination Of Urinary Microalbumin

    Significance Of Determination Of Urinary Microalbumin

    August 8, 20181. Early predictive value of renal damage in hypertensionThe urinary albumin of hypertensive patients is significantly higher than that of normal people, which is positively correlated with the increa...view
  • TEL:+86-21-68129666
  • FAX:+86-21-68129222
  • EMAIL:international@chemtronbio.com
  • ADDRESS:No.518, Qingdai Rd., International Medical Park, Pudong, 201318, Shanghai, P.R.China